Follow these procedures for the operation of the unit.
•
Power up and initial indication.
•
Setting the temperature control.
•
Setting the limit control.
•
Emptying and cleaning the unit.
Follow these steps to turn On the unit to prepare it for use:
1. With the unit connected to an appropriate power source, set the power
switch to the ON position.
2. The green indicator will illuminate when the unit is applying heat to
the bath. This indicator will cycle on and off during normal operation.
Follow these steps to set the Temperature Control so that it maintains the
desired temperature.
1. Turn Limit Control knob fully clockwise.
2. Turn the Temperature Control to a setting which approximates the
desired temperature. Measure the bath temperature with the supplied
thermometer or other temperature measuring device. Allow the bath
temperature to stabilize, then readjust the control knob as necessary to
achieve the proper temperature.
3. Set the Limit Control according to the procedure below.
Note The safety control is not as accurate a control as digital control.
Note The thermometer contains kerosene and is not expected to be a
health hazard in the normal intended use of the thermometer.