Reference Manual
00809-0400-4004
7
Advanced configuration
Advanced configuration options are used to configure the flow meter for a wider range of
applications and special situations.
7.1
LCD display
ProLink III
The LCD display (option M5) provides local indication of the output and abbreviated
diagnostic messages governing operation of the flow meter. You can select any of the
following variables to be displayed, where at least one must be selected:
•
Primary Variable
•
Percent of Range
•
Totalizer Value
•
Shedding Frequency
•
Mass Flow
•
Velocity Flow
•
Volume Flow
•
Process Temperature
•
Pulse Frequency
•
Shedding Frequency
•
Electronics Temperature
•
Signal Strength
•
Corrected Volume Flow
Note
Analog Output and % of Range can be selected, but provide no useful information when
Modbus protocol is used.
7.2
Compensated K-factor
ProLink III
The compensated K-factor is based on the reference K-factor as compensated for the
given process temperature, wetted materials, body number, and pipe ID. Compensated K-
factor is an informational variable that is calculated by the electronics of the flow meter.
The reference K-factor is factory set and is displayed on the support tube label. The
reference K-factor should only be re-configured in the device in the event of transmitter
replacement. Contact technical support for details.
Reference Manual
Device Tools → Configuration → Display Variables
Device Tools → Configuration → Device Setup
Advanced configuration
October 2021
47