4
Operation
Select restrictors
Start up the spreadsheet "splitter3_calc_rev1.xls" in Excel. Select the "3-Way
(Makeup)" worksheet. We will choose to have 4 mL/min go to the MSD initially.
This flow is acceptable with a performance turbo system. Fill in the input
column as shown
as Detector 2, and the ECD as Detector 3.
Figure 18
The Effluent Splitter calculator
42
(Figure
18) with the MSD assumed to be Detector 1, the FPD
Installation and Operation Guide