How It Works

The splitter divides the effluent from a column among three different
detectors. The detectors can be operating at different pressures, that is, any
mix of the following can be used:
• Atmospheric pressure
• Below atmospheric pressure
• Above atmospheric pressure
The split ratio is determined by the length and diameter of tubing connecting
the splitter to the detectors. Tubing dimensions may be determined from
Table 2
included for calculating tubing dimensions for special situations.
Figure 1 shows the plumbing configuration for the G3183B splitter.
Figure 1
Installation and Operation Guide
FID (flame ionization detector)
TCD (thermal conductivity detector)
NPD (nitrogen phosphorus detector)
ECD (electron capture detector)
FPD (flame photometric detector)
MSD (mass selective detector)
AED (atomic emission detector)
on page 26 in this manual or from a spreadsheet calculator that is
Inlet
Aux EPC
Column
GC oven
Splitter plumbing
Det 1
Det 2
Restrictor 1
Restrictor 2
Restrictor 3