Select And Install A Restrictor - Agilent Technologies QuickSwap G3185B Installation And Setup

Table of Contents

Advertisement

5 For users who plan to use a CI source (CI MSDs only), check the MSD

2. Select and Install a Restrictor

1 Install the lock nut on the end of the MSD transfer line
Figure 6
2 Restrictors are precut lengths of deactivated fused-silica tubing with
QuickSwap Installation and Startup
ChemStation version.
If using MSD ChemStation version D.01.xx or earlier, install the EI source.
Agilent recommends using EI mode for easier installation and leak testing
in this case.
If using MSD ChemStation version D.02.xx or later, proceed with either the
EI or CI source installed.
See also
"Final leak test procedure for CI sources"
threads facing the MSD. Run it down until it touches the transfer line (stops
turning). Do not tighten it.
Transfer line
Lock nut
Threads engaged
Installing the lock nut
swaged ferrules attached.
For initial installation, Agilent recommends installing a 100-µm
restrictor for use with the checkout methods supplied on the QuickSwap
User Information CD.
If not using one of these methods, or if this is not an initial installation,
refer to the
Reference Manual
on page 35.
Touching transfer line
to choose an appropriate restrictor.
Overview
(Figure
6) with the
19