compact piezo LeD
Rinse the internal duct of the insert tip
with demineralized water at a pressure of
3.8 bar for at least 10 seconds in order to
eliminate any residual dirt;
Place the insert tip in the ultrasonic tank,
submerged in the enzymatic detergent
solution at 24 ° c ± 2 ° c and run a cycle
of at least 20 minutes;
Remove the insert tip/s from the
ultrasonic cleaning tank and rinse it in
demineralized water;
Brush the internal and external surfaces
of the insert tip using a clean brush with
soft nylon bristles, under running water;
22
16
x10 sec.
H₂O
17
18
19