COMPACT PIEZO P2K CONTENTS Introduction Intended Use Description of the Appliance 1.2.1 Eligible Patient Group 1.2.2 Patient Selection Criteria 1.2.3 Indications for Use 1.2.4 Users Disclaimer Safety Requirements Symbols Identification Data Appliance ID Plate 2.2 Scaler Handpiece Identification Data 2.3 Inserts Identification Data Delivery List of Components...
Page 5
7.1 Electromagnetic Compatibility IEC/EN 60601-1-2 7.1.1 Guide and Manufacturer’s Declaration - Electromagnetic Emissions 7.1.2 Accessible Parts of the Casing 7.1.3 Guide and the Manufacturer's Declaration - Electromagnetic Immunity 7.1.3.1 Power Connection BC Input 7.1.3.2 Points of Contact with the Patient 7.1.3.3 Parts Accessible to the Input / Output Signals 7.1.4 Specifications of the Tests for the Immunity of the Accessible Parts of the Casing to the Wireless RF Communications Device Troubleshooting Quick Troubleshooting 8.2 Shipping to an Authorised Mectron Service Centre Warranty...
Page 6
COMPACT PIEZO P2K PAGE LEFT INTENTIONALLY BLANK...
The Mectron piezoelectric ultrasonic scaler devices can be used on patients of any age and gender requiring a tooth polishing treatment. There are no contraindications for specific population segments. WARNING: The appliance must be used...
Failure to observe this regulation can cause serious injury to the patient, the operator, and damage/failure of the device. Description of the Appliance COMPACT PIEZO P2K is a multi-purpose levels. piezoelectric ultrasonic scaler. The appliance is equipped with an automatic It is designed to equip the operator with a...
• Use of non original MECTRON inserts The manufacturer MECTRON cannot be held resulting in permanent damage to the liable, expressly or implicitly, for any type of...
An electrosurgical scalpel or other At first use and after each treatment they electrosurgical units positioned near the must be cleaned and sterilised by carefully COMPACT PIEZO P2K appliance may interfere following the instructions in Chapter 6 on with the correct operation of the appliance page 19 of the Use and Maintenance itself.
INTRODUCTION Symbols Symbol Description Symbol Description Device compliant Nemko brand with Regulation (EU) Compliance with UL - 2017/745. Notified 0051 CSA regulations body: IMQ S.p.A. Caution! Read the Medical device instructions for use Operating instructions Manufacturer Date of manufacture Serial number Batch number Product code The sterilisable...
Mectron S.p.A. Via Loreto 15/a number. The ID plate is located on the 16042 Carasco -GE- Italy COMPACT PIEZO P2K side of the COMPACT PIEZO P2K module. 50/60 Hz / 32V Min. 40 VA The complete technical specifications are 04000079 YYYY-MM-DD provided in...
Chapter 1.5 on page 5 03120016 0051 YYYY-MM SCALER ABLATORE 102000000 Bar Code SCALER HIBC 128 03120016 HANDPIECE Inserts Identification Data The name of the insert itself (Reference 1), MECTRON logo (Reference 2), and insert batch number (Reference 3) are laser engraved on each insert.
DELIVERY List of Components Figure 1 on page 11. can be ordered separately (see Table 4 on COMPACT PIEZO P2K is supplied with the page 11). standard equipment (see Table 2 on page NOTE: Both the items included in the 8), a set of variable accessories depending...
Page 15
DELIVERY Accessories available for order with standard supply Item Code Description Ref. 03570004 “ICS” series reusable base insert 03590009 "IC1" tip 02900112 tips kit (5 pcs.) DB1 reusable base support for 03570001 inserts 03590001 TA12D60 diamond insert 03590004 TA12D90 diamond insert 03590002 TA14D60 diamond insert 03590005 TA14D90 diamond insert 03590007 TA14D120 diamond insert 03590003 TA16D60 diamond insert...
Page 16
COMPACT PIEZO P2K Accessories available for order with standard supply Item Code Description Ref. 01560022 Layout A 01560023 Layout B 01560024 Layout C 01560026 Layout D Wiring 01560027 Layout E 01560031 Layout F 01560101 Layout G 01560141 Layout H 02150174 IT version 02150175 EN version Use and Maintenance Manual...
Page 17
Keep the packaging in case any items need to storage must be carried out with particular be sent to an Authorised MECTRON Service care. Centre and to store the appliance during long All the material sent by MECTRON was periods of non-use.
The dental unit to which COMPACT PIEZO P2K essential performance. is connected must be a medical device that complies with general and applicable product CAUTION: Check that the dental unit to...
Figure 2 – Dimensions of the module. Electrical Connections COMPACT PIEZO P2K allows various electrical electrical connections, configurations to satisfy the characteristics of COMPACT PIEZO P2K device is equipped with the dental unit in which it is housed. 2 connectors (see Figure 3 on page 14): dental units which •...
Page 20
COMPACT PIEZO P2K 16 15 14 13 12 11 10 9 Figure 3 – Module connectors. M1 Connector Pin-Out Pin Signal Signal Reserved Bypass capacity Pedal + Power Selection Valve AC In Power Control AC/DC In Pedal - +Vdc Reserved Table 5 – M1 Connector Pin-Out.
INSTALLATION 4.3.1 Alternate 24V Power Supply To power the module at 24V~ 50/60 Hz connect pins 13 and 14 (see Figure 3 on page 14 and Table 5 on page 14). Connect pins 8 and 9 to insert the bypass capacity present on the board to filter any disturbances on the power supply. Observe the voltage and current values indicated on the plate. 4.3.2 Direct 32V Power Supply To power the module at 32V connect the power supply positive on pin 14 and...
COMPACT PIEZO P2K 4.3.3.2 Power Regulation with 10 kOhm Potentiometer Connect the 10 kOhm potentiometer between pins 6 and 10 and a 20 kOhm, 10 kOhm linear 1/8W, 1% resistance between pins 12 and potentiometer The maximum power corresponds to resistance 0. 20 kOhm 1/8W, 1% resistance 4.3.3.3 Power Regulation with 4.7 kOhm Potentiometer Connect the 4.7 kOhm potentiometer between pins 6 and 10 and a 39 kOhm, 4.7 kOhm linear...
12 and 10. 0-10V 1mA 10 kOhm 1/8W, 1% resistance 4.3.4 Solenoid Valve Control The COMPACT PIEZO P2K provides an enabling in output for a solenoid valve Solenoid valve (optional) on the handpiece irrigation circuit. The 24V 2,5W output distributes 24V with maximum power 2.5W.
COMPACT PIEZO P2K 4.3.5.2 Enabling of Pedal with Positive Signal The pedal must be connected to pins 8 and 11. Pedal 4.3.6 Other Connections Pins 1 and 16 must always be connected to each other.
Standard Connections facilitate installation NOTE*: Near the connection on the module COMPACT PIEZO P2K in the different types of it is advisable to wind the power line 3 times dental units, there are 7 available “standard” on a ferrite (see following chapters). The connection layouts, each equipped with a ferrite is not supplied standard.
COMPACT PIEZO P2K 4.3.8.1 Layout A The power supply (24 V~ 50/60 Hz ±10% or 32 V ±10%) and consent are provided by the pedal. The power is regulated by a 2.2KΩ linear potentiometer (see Figure 5 on page 20). Fuse holder, fuse and pedal are not included in the Mectron supply 2.2 KΩ...
Minimum 40VA Negative Fuse holder, fuse, instrument Solenoid valve tray and pedal are not (optional) included in the Mectron supply 24 V 2,5W maximum Scaler handpiece cord Green 8 7 6 5 4 3 2 1 Figure 7 – Layout C 4.3.8.4...
COMPACT PIEZO P2K 4.3.8.5 Layout E The power supply (24 V~ 50/60 Hz ±10% or 32 V ±10%) and consent are provided by the pedal. The power is regulated by a 4.7KΩ linear potentiometer (see Figure 9 on page 22). Fuse holder, fuse and pedal are not included in the Mectron supply 2.2 KΩ linear...
The power is regulated by a variable voltage between 0 and 5 V (see Figure 11 on page 23). Fuse holder, fuse and pedal are not included in the Mectron supply Power supply 24 Vac ± 10% 50/60 Hz Pedal...
COMPACT PIEZO P2K Water Connection The COMPACT PIEZO P2K requires a water and solenoid valve, be fitted with supply line to cool the handpiece and inserts a valve dedicated to supplying the during the operating cycle. COMPACT PIEZO P2K handpiece and The device is sold with a connecting cable...
DIAGNOSTICS DIAGNOSTICS The appliance is equipped with an APC components, and a two-coloured diagnostic automatic protection circuit which LED (green and red), positioned on the board, activated when operating faults occur on the which indicates when the module is powered. Description of Diagnostic Warnings 5.1.1 Appliance Powered Without Pedal Consent...
• Inserts, when worn or broken; injuries. • Inserts tightening wrench, when worn or COMPACT PIEZO P2K must be disposed of and broken. treated as waste subject to separate collection. Disposable materials and materials that carry...
30 sec. ON - 120 sec. OFF without irrigation Power adapter compliant with IEC/EN 60601-1: Power Supply • 24 V~ 50/60 Hz or • 32 V Max. Power Consumption 40 VA Fuse T 2AL, 250V (not included in Mectron supply) Automatic Scan Working Frequency From 24 KHz to 36 KHz Adjustable depending type Power installation.
Guide and Manufacturer’s Declaration - Electromagnetic Emissions COMPACT PIEZO P2K is designed to operate in the electromagnetic environment specified below. The purchaser or user of COMPACT PIEZO P2K should ensure that it is used in such an environment. Emissions Test...
TECHNICAL DATA 7.1.2 Accessible Parts of the Casing COMPACT PIEZO P2K is designed to operate in the electromagnetic environment specified below. The purchaser or user of COMPACT PIEZO P2K should ensure that it is used in such an environment. Essential EMC Electromagnetic Phenomenon standard or test Immunity test values...
Guide and the Manufacturer's Declaration - Electromagnetic Immunity 7.1.3.1 Power Connection BC Input COMPACT PIEZO P2K is designed to operate in the electromagnetic environment specified below. The purchaser or user of COMPACT PIEZO P2K should ensure that it is used in such an environment. Essential EMC Electromagnetic Phenomenon standard or test...
Page 37
(common mode) and at ± 1 kV between line(s) voltage values. and line(s) (differential mode). b) During the test, all COMPACT PIEZO P2K cables must k) Not applicable to CLASS II devices and systems. be connected. l) It is necessary to use direct coupling.
COMPACT PIEZO P2K 7.1.3.2 Points of Contact with the Patient COMPACT PIEZO P2K is designed to operate in the electromagnetic environment specified below. The purchaser or user of COMPACT PIEZO P2K should ensure that it is used in such an environment. Essential EMC Electromagnetic Phenomenon standard or test...
TECHNICAL DATA 7.1.3.3 Parts Accessible to the Input / Output Signals COMPACT PIEZO P2K is designed to operate in the electromagnetic environment specified below. The purchaser or user of COMPACT PIEZO P2K should ensure that it is used in such an environment. Essential EMC Electromagnetic Phenomenon standard or test Immunity test values...
Parts of the Casing to the Wireless RF Communications Device COMPACT PIEZO P2K is designed to operate in an electromagnetic environment in which radiated RF disturbances are under control. The purchaser or operator of COMPACT PIEZO P2K can help prevent electromagnetic interferences by guaranteeing a minimum distance between the mobile and portable RF communication devices (transmitters) and COMPACT PIEZO P2K, as recommended below, in relation to the maximum output power of the radio communication devices.
WARNING: Portable RF communication level, the distance between the transmitter equipment (including peripheral devices such antenna and COMPACT PIEZO P2K can be as antenna cables and external antennas) reduced to 1 m. The test distance of 1 m is must not be used closer than 30 cm to any allowed by IEC 61000-4-3.
Page 42
COMPACT PIEZO P2K Problem Possible Cause Solution The insert is the type without Use an insert type with a a liquid passage liquid passage The Water function is not Activate the Water function active Unscrew the insert from the handpiece and release the...
• Details of owner and contact number; accessories. • Product name; Clean and sterilise all parts that need to be • Serial number and/or batch number; sent to an Authorised Mectron Service Centre, following the instructions in Chapter 6 on page • Reason for return / description of failure; of the Use and Maintenance Manual.
• Damage due to incorrect use or Mectron provides warranty carelessness, or due to connection to a COMPACT PIEZO P2K, purchased new from voltage other than that envisaged. a Mectron dealer or importer, which covers • Expired warranty. defects in material and workmanship for a...
Page 45
In order to avail of the warranty service, the client must return the appliance, at their own expense, to the MECTRON dealer/importer from which they purchased the product. The appliance must be returned together with the original packaging, accompanied by all the accessories and by a form containing: •...