Remove the torque wrench from the
ultrasonic cleaning tank and rinse it
under running water to remove any
residual detergent.
Brush the internal and external surfaces
of the torque wrench with a clean brush
with soft nylon bristles, under running
water.
Process validated by independent bodies with enzymatic detergent.
6.2
Cleaning Check
6.2.1
Required Materials
• Light source;
• magnifying glass 2,5X.
once the cleaning operations have been
completed, check the scaler handpiece
and scaler nose cone under an adequate
source of light, if necessary using a
magnifying glass 2.5X, paying attention
to any parts that may conceal soil residue
(threading, cavities, grooves) and, if need
be, repeat the cleaning cycle if soil is
still visible. Finally, check the integrity
of those parts and those elements that
could have deteriorated during use;
CLEANING AND STERILIzATION
26
1
25
Need help?
Do you have a question about the compact piezo LED and is the answer not in the manual?
Questions and answers