Rinse the insert tip in demineralized
water;
Rinse the internal duct of the insert tip
with demineralized water at a pressure of
3.8 bar for at least 10 seconds in order to
eliminate any dirt residues.
Process validated by independent bodies with enzymatic detergent.
6.1.5
Torque Wrench
Place the torque wrench horizontally
in a clean container. add enough of
the enzymatic detergent solution to
completely cover the torque wrench.
Soak the torque wrench in the enzymatic
detergent solution for 10 minutes at
room temperature (22 ° C ± 2 ° C).
CLEANING AND STERILIzATION
20
21
x10 sec.
H₂O
22
23