6.3
Drying and Lubrication
6.3.1
Necessary Materials
• compressed air;
• Soft cloth with low fibre release;
• medical grade lubricant.
thoroughly dry all parts of the scaler
handpiece, scaler nose cone, and light
pipe by blowing them with compressed
air;
CAUTION: the scaler handpiece
electrical contacts must be dry at the
start and end of the sterilisation cycle,
before the cord is connected to the
appliance. Always make sure that the
electrical contacts of the connector
are perfectly dry, if necessary dry them
blowing with compressed air.
CAUTION: Before starting the
sterilization cycle, make sure that
the insert tip is thoroughly dry both
internally and externally. to do this,
blow compressed air both externally and
through the internal passage hole. this
will prevent the appearance of stains,
streaks on the surface or oxidation inside
the insert tip.
CLEANING AND STERILIzATION
NOTE: Adopt a medical-grade lubricant
approved by FDA, for example, the NSK Pana-
Spray or similar.