During the immersion in the enzymatic
solution, gently brush all surfaces to
remove any visible dirt.
Use a clean, soft nylon bristle brush for
external surfaces, a clean, soft nylon
bristle brush for internal cavities and
crevices.
NOTE: carefully brush all the following
parts of the insert tip for about 20
seconds:
• through holes and internal ducts;
• Difficult to clean areas such as
sharp edges and in particular the
interstices between the cutting
cusps;
• internal cavities, grooves and
crevices.
Remove the insert tip from the enzymatic
detergent solution.
thoroughly rinse and brush all surfaces of
the insert tip (see previous point) under
running water for at least 10 minutes;
Use a disposable syringe to aspirate and
inject the enzymatic detergent solution
into hard-to-reach areas (through holes
/ cannulas). Repeat this operation three
times to ensure the effective removal
of dirt from the internal surfaces of the
through hole;