Instructions for Use
Surgical Laser Unit Limax
6.2.4
MCB Socket Limax
6.2.5
Accessory Port
6.2.6
Interlock/Warning Connector
Fig. 6-3:
Solder side of the 7-pin plug Interlock/Warning
Revision 5
®
120
®
120
Using the MCB cable, connect the MCB socket of the
®
Limax
120 (4, Fig. 6-2) to the MCB socket of the
integrated VAC, see section 6.2.16 "VAC MCB Socket",
page 35.
Possibility to connect external devices (6, Fig. 6-2) and
control them via the foot switch. Connecting cables are
available from the Martin Service Center on request, see
section 2.2 "Hotline", page 11.
The foot switch operates a floating switching contact that
can be used for controlling accessory components.
Connector for interlock/warning plug (7, Fig. 6-2):
Interlock/Warning
The connection has 2 functions:
Laser warning lamp contact
•
Door interlock
•
Laser warning lamp contact
This contact (1) activates the warning lamps when
switching on the device (key switch on I).
All accesses to the laser area should be equipped with laser
warning lamps.
The potential-free switch contact (max. 48 V, 1 A) can
•
be used via the supplied red interlock plug with bypass
(REF 08-507-00-09).
29