Gemini VII 2390 Service

Manual Control

Some components of the analyzer can be manually controlled using commands provided in the
software.

Embedded

An overlay is used when manually controlling the analyzer with a keypad. The overlay is
placed on the keypad to guide you through the functions.
RAW/TRUE
B/S
Feb 09
Displays the raw A/D readings for the sample, balance,
and Po transducers or quantity adsorbed transducer, or
the true pressure and volume readings.
Disables/enables the balance servo.
• A B in the display window indicates that the balance
servo and sample servo are enabled.
• An S indicates that only the sample servo is enabled;
the balance servo is disabled.
For this command, you must press the
pressing
.
8
Manual Control
key before
Alt
2-15