Process Errors Without Messages - Endress+Hauser Proline Promass 83 Operating Instructions Manual

Coriolis mass flow measuring system
Hide thumbs Also See for Proline Promass 83:
Table of Contents

Advertisement

9 Trouble-shooting
82
9.3

Process errors without messages

Symptoms
Remark:
You may have to change or correct certain settings in parameters in order to rectify the fault. The parameters listed
below, e.g. "Install. Direction Sensor" are described in detail in the "Description of Device Functions" manual.
Measured value reading fluctua-
tes even though flow is steady.
Measured value reading shown
on display, even though the fluid
is at a standstill and the measu-
ring tube is full.
The fault cannot be rectified or
some other fault not described
above has occurred.
In these instances, please contact
your Endress+Hauser service
organisation.
Proline Promass 83 FOUNDATION Fieldbus
Rectification
1. Check the fluid for presence of gas bubbles.
2. Increase the following value:
– FF: Analog Input function block
– Local display: HOME
BASIC FUNCTION
CONFIGURATION
FLOW DAMPING
3. Increase the value for display damping:
– FF: Transducer Block
Parameter "User Interface - Damping"
– Local display: HOME
USER INTERFACE
FIG.
DISPLAY DAMPING
1. Check the fluid for presence of gas bubbles.
2. Enter or increase the value for the switching point of low flow cutoff:
– FF: Transducer-Block
Parameter "Low Flow Cutoff - On-Value"
– Local display: HOME
BASIC FUNCTION
CONFIGURATION
ON-VAL. LF-CUTOFF
The following options are available for tackling problems of this nature:
Request the services of an Endress+Hauser service technician
If you contact our service organisation to have a service technician sent out, please
be ready with the following information:
– Brief description of the fault
– Nameplate specifications (Page 9 ff.): Order code and serial number
Returning devices to Endress+Hauser
The procedures on Page 8 must be carried out before you return a flowmeter requi-
ring repair or calibration to Endress+Hauser.
Always enclose a duly completed "Declaration of contamination" form with the
flowmeter. You will find a preprinted "Declaration of contamination" form at the
back of this manual.
Replace transmitter electronics
Components in the measuring electronics defective
order replacement
Page 83
Parameter PV_FTIME
SYSTEMPARAMETER
CONTROL
BASIC CON-
PROCESSPARAMETER
Endress+Hauser

Advertisement

Table of Contents
loading

Table of Contents