Installing the PREVI™Color Gram

Installation procedure

I n s t a l l a t i o n p r o c e d u r e
CAUTION!
I n s t a l l i n g w a s t e t u b i n g
Clip the waste tubing
into the PREVI™Color
Gram and push until
you hear a "click".
Adjust the tube length
(L < 2m).
Connect the waste
tubing to the waste
container.
Drip tray
Install the waste
container into a drip
tray to prevent from
spilling any liquid.
3-2
Ensure that the distance between the back panel of the PREVI™Color Gram and
the wall is at least 30 cm (12 inches).
Before the PREVI™ Color Gram is installed and switched ON it should be left to
stand for at least three hours so that there is no possibility of condensation
causing a hazardous situation.
Keep the waste container tube straight and as short as possible:
2 meters (79 inches) maximum.
The waste container tube should be positioned at a lower level than the table.
If possible, bioMérieux recommends you to install the instrument at an extremity
of the bench top.
PREVI™ Color Gram User's Manual