Maintenance and cleaning
Nozzle handling
Reassemble the
nozzle by inserting
the swirl cone into the
compression screw.
Hold all the parts in a
vertical position until
reassembly is
completed.
Return the nozzle to
its original location.
Repeat instructions
1 to 8 for each nozzle.
Prime each reagent
channel.
Perform pattern and
volume tests before
reusing the
instrument.
IMPORTANT!
If the test is still not correct, see "Unclogging the tubing", page 5-8.
PREVI™ Color Gram User's Manual
5-7