United Kingdom Conformity Assessed
Electromagnetic Compatibility Regulations 2016
The Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment Regulations 2012
Supply of Machinery (Safety) Regulations 2008
IMPORTANT
Any serious incident that has occurred in relation to this product should be reported
to Miltenyi Biotec B.V. & Co. KG using the contact information provided.
4.2 Technical data
Safety and performance of the instrument may be compromised if the
CliniMACS Formulation Unit is used outside its specifications. Do not use
the instrument outside its specifications.
Before connecting the instrument to the CliniMACS Prodigy, the
configuration of the CliniMACS Prodigy must be checked and adjusted if
necessary. Contact authorized local Miltenyi Biotec Service Provider or
Miltenyi Biotec Technical Support for installation of the instrument.
Technical data
Model
CliniMACS Formulation Unit (REF 170-075-703)
Dimensions
Width: 62.4 cm housing
Depth: 9 cm housing (without mounting guide)
Height: 8.5 cm housing (without valves)
Weight
Approx. 9 kg (excluding the weight of attached consumables)
Emission sound
<79 dB(A)
pressure level
Input voltage
24 V DC 5.5 A max (supply by CliniMACS Prodigy)
Operation
Temperature: +15 °C to +25 °C (+59 °F to +77 °F)
conditions
Humidity: 10% to 75% relative humidity, non-condensing
Altitude: <2,000 meters above sea level
Storage conditions
Room temperature
Avoid condensing conditions.
Positioning
Installed on the bag compartment of the CliniMACS Prodigy
CliniMACS Formulation Unit and CliniMACS Prodigy must have
a distance of at least 14 cm to walls and other obstacles.
Table 4.1: Technical data of the CliniMACS Formulation Unit
4
21