ERC/AHA Guideline 2010

5.2 Cardio pulmonary resuscitation

After the analysis or shock, application the device requests cardio pulmonary resuscitation.
This consists of five (seven) CPR cycles consisting of 30 chest compressions and two breaths.
The necessary actions will be announced by corresponding voice messages in the first cycle:
< Cardio pulmonary resuscitation >
< Give 30 chest compressions > (< Give 15 chest compressions >)
< Give 2 rescue breaths >
After the spoken message, 30 bleeps will follow during the chest compressions at the
recommended frequency of 100 compressions / minute. The rescue breaths are also
supported by an appropriate sound signal. From the second to fifth CPR cycle, only these
sound signals are emitted.
5.3 Message < Check electrodes >
The instructor can simulate the waste of an electrode by pressing the  button on the remote
control.
In this case, the device repeats the following voice messages:
< Check electrodes >
< Apply electrodes one after the other to patient's bare chest >
If the plug on the PRIMEDIC™ SavePads Trainer has not been inserted in the unit, the
following additional instruction appears
< Plug in electrode cable >
These voice instructions to be repeated for the duration of one minute. If the device has not re-
signalled a good electrode contact before this time, there are instructions for five (seven)
cycles for cardiopulmonary resuscitation:
< Give 30 chest compressions > (< Give 15 chest compressions >)
< Give 2 rescue breaths >
Afterwards, the device will once again give instructions to attach the electrodes for a maximum
period of one minute. This process is continued until the device signals a good electrode
contact and begins the rhythm analysis.
20 / 24
23560 HeartSave AED Trainer / GB / B02
METRAX GmbH