ERC/AHA Guideline 2010

5 ERC/AHA Guideline 2010

5.1 Voice output of the device

After the self-test has been successfully performed by the device, the following BLS-language
instruction (BLS = Basic Life Support) is issued:
<Adult mode>
< Call emergency services >
After that the following voice prompt is issued for a period of one minute:
< Apply electrodes one after the other to patient's bare chest >
If the Electrode plug is not yet inserted in the device, the audio prompt will be issued for one
minute:
< Apply electrodes one after the other to patient's bare chest >
< Plug in electrode cable >
If, up to this point the device does not detect any patient impedance, because the  button on
the remote is not pressed yet, there are instructions for one cycle for cardiopulmonary
resuscitation (CPR):
< Give 30 chest compressions >
< Give 2 rescue breaths >
Afterwards the device will give instructions to attach the electrodes for maximum one minute.
This process is continued until the device, a first electrode pad (corresponding to the actual
case of a valid patient impedance) is signalled. By pressing the  button on the remote
control.
< Analysing rhythm, Do not touch the patient >

5.1.1 Defibrillation necessary:

< Shock advised >
< chest compressions > (voice message omitted in Pediatric Mode)
< Stand clear of patient >
< Press lit shock button now >

5.1.2 Defibrillation is not necessary:

< No shock advised >