FujiFilm PL-G.MAX Operation Manual page 5

Pipeline and cable locator
Table of Contents

Advertisement

Troubleshooting known errors that may occur
"Push Down" is displayed when the receiver is at center of the line and the depth button pressed. This may indicate
the footplate assembly has unthreaded, so check and screw it tight if it has become loose. Another possible cause is
due to a dirty optic sensor. This will prevent the receiver from being able to read the position of the slide pipe
"Over Current" is displayed on the transmitter. This issue may stem from multiple issues, but this transmitter is not
working to manufacturer specs.
Difficult to extend and depress slide pipe. The slide pipe should move freely, with minimal resistance and noise. Be
sure to continually clean slide pipe of moisture and debris, to keep it from jamming up and bringing dust, dirt and water
inside the housing of the receiver.
Acid Damage. When changing batteries always check that battery compartments are free of dirt, moisture and acid
damage. All batteries and equipment should be taken inside during cold temperatures and you should never heat up
cold batteries. Monitoring of the battery housing for traces of acid damage can help avoid serious damage, if an issue
arises see below for "leaking Battery Procedure"
Weak Signal. This issue can occur in the transmitter or receiver and will be noticed during the testing by comparing it
to previous values recorded. If a drastic change is noticed in location, depth or current measurement or sensitivity are
found, the locate tool should be tagged and sent in for service / calibration.
If any of the above problems occur ensure that the locate tool is tagged and sent in for service/calibration.
Leaking Battery Procedure
Whenever replacing batteries, thoroughly examine dead or low batteries for evidence of leakage. Look in battery
compartments and look for evidence of residue. Do not put new batteries in transmitter or receiver if any residue is
observed. In the event of leaking batteries, remove them to reduce further damage.
If an incident has occurred:
1. Remove batteries. Do not replace with new batteries.
2. Neutralize and clan battery compartment with baking soda, 3-4 tablespoons in battery compartment.
3. Close battery compartment and shake
4. Open battery compartment shake out any excess baking soda and clean battery terminals with soft brush
5. The locate tool should be sent in for service/calibration after this procedure is complete.
Wet Locate Tool Procedure
Receiver
1. Remove the battery pack
2. Extend slide pipe fully out and towel off any moisture
3. Lay receiver on its side, in a warm dry environment
4. Do not place in close proximity of a heat source
5. Do not stand it right side up, or upside down, as any condensation may migrate to the circuit boards in the
head or the top of the receiver
Transmitter:
1. Remove all batteries and towel out any moisture in the battery compartment or anywhere else on transmitter.
2. Lay transmitter flat with battery compartment open to dry, in warm dry environment.
3. Do not place in close proximity to a heat source
4. The locate tool should be tested after drying is complete
Repair and Calibration Protocol
Prior to sending tool in for service and or calibration, call and explain issue to Fuji Pipe Locators Ltd
(403) 277-3300 or ben@leaklocate.com
- 3 -

Advertisement

Table of Contents
loading

Table of Contents