Pull Up Measurement - FujiFilm PL-G.MAX Operation Manual

Pipeline and cable locator
Table of Contents

Advertisement

PULL UP
83kHz
SENS.30
2.3
m
Measured depth value
OPERATION OF RECEIVER
Figure 1
Measured value
is displayed
055
Current index value
5-2-4 Pull-up Depth Measurement
Set up the transmitter
☞ Orient the Transmitter correctly to the pipeline
Operation of Receiver (Standard Depth Measurement)
Enter MAXIMUM MODE
*Set as Sonde Mode when in use of Sonde
Orient the receiver above the pipeline as shown by the arrows
in figure 1, and stop when the LCD screen reaches maximum
indication. See page s6 (section 5-2-2)
Push down the antenna "completely" and press the depth
measurement switch①.
NOTE: depth cannot be measured with Bar mode.
When you see "PULL UP" on the LCD screen, fully pull
Up/extend the antenna. ②
After pulling up completely, ③ the measured value will be
displayed on the LCD screen.
【Attention】
・Do not move the antenna after pressing the depth
measurement key until "PULL UP " appears on the display.
・Do not move the antenna after pulling up the antenna until
the depth result is displayed.
・After taking a depth measurement, pressing the Sensitivity
Adjustment key will display the previous screen. Pressing the
depth key will display the horizontal displacement depth
measurement screen.
- 28 -