3.2.9 Salinity Value

Enter the sample salinity value as 0.0 to 50.0 ppt. The
entered salinity value is used for salinity corrected
dissolved oxygen measurements.
Press the  or  key to scroll through the Setup
Menu list until the Salinity Value item is shown.
Press the  or  key to set the salinity value.
Press the  key to increase the value by 0.1. Press
and hold the  key to quickly increase.
Press the  key to decrease the value by 0.1. Press
and hold the  key to quickly decrease.
Press the "Enter" key to save the change.

3.2.10 Read Type

Set the Read Type as Continuous, Auto-Read or Timed to
define how measurements are performed and when
measurements are saved and exported.
Continuous:
In the Continuous measure mode, measurements are
continuously updated on the display and the stopwatch
(stabilizing) or checkmark (stable) icon indicates the
measurement stability status. Press the "Log/Export" key
to save a measurement to the data log and export.
Auto-Read:
In the Auto-Read measure mode, press the "Measure" key
to start a measurement. When the measurement is stable,
the checkmark (stable) icon is shown and the
measurement is locked on the display until the "Measure"
key is pressed again. The stable measurement is
automatically saved to the data log and exported.
Timed:
In the Timed measure mode, the measurement values are
continuously updated on the display. Measurements are
automatically saved to the data log and exported at the
selected time intervals, from 5 seconds to 60 minutes, the
entire time the meter is in the measurement mode.
Press the  or  key to scroll through the Setup
Menu list until the Read Type item is shown.
Press the  or  key to set the read type.
Press the "Enter" key to save the change.
Eutech DO 1730 Dissolved Oxygen Meter
19