Set Value Reading In The General-Purpose Data Area (Using Request Flag 2) - Unipulse F372A Operation Manual

Cc-link i/f
Hide thumbs Also See for F372A:
Table of Contents

Advertisement

8.Setting Procedure
ex.) When 4 stations occupied (writing in Setting value)
Request flag 2
Request flag 2 response
General purpose data area
(RWw000C)
(RWw000D)
(RWw000F)
Command No.
(RWw000E)
R/W response
Command No. response
■Set value reading in the general-purpose data area
(using request flag 2)
Turn ON R/W for reading out Setting value.
The WORK data of selected WORK No. is output.
Read out General purpose data area (response) after confirming Request flag 2 response ON.
① Input Work No. to read to the Work No. area.
② Input the command of the set value to be read in the Command area in the general-purpose data area.
③ Set R/W of the remote output to 1.
④ Make request flag 2 of the remote output 0→1.
⑤ After confirming response flag 2 of the remote input 0→1, make request flag 2 of the remote output 1→0.
⑥ Responses are stored in R/W (response) of the remote input and general-purpose data area (response).
Confirm the contents of the general-purpose data area (response).
ex.) When 4 stations occupied (Reading out Setting value)
Request flag 2
Request flag 2 response
WORK No.
(RWw000F)
Command No.
(RWw000E)
R/W response
Command No. response
General purpose data area
(RWr000C)
(RWr000D)
18
(RY0002)
(RX0002)
R/W
(RY0003)
WORK No.
(RX0003)
(RWr000E)
(The Upper row of signal level is ON. The Lower row of signal level is OFF.)
(RY0002)
(RX0002)
R/W
(RY0003)
(RX0003)
(RWr000E)
response
(The Upper row of signal level is ON. The Lower row of signal level is OFF.)
Range: 0 to 15