Switching On The Unit; System Test - KLS Martin group LIMAX 120 Operating Instructions Manual

Surgical laser unit
Hide thumbs Also See for LIMAX 120:
Table of Contents

Advertisement

6.8

Switching on the Unit

6.9

System Test

50
Surgical Laser Unit Limax
The red emergency stop switch must be in
pulled-out position.
Insert key, then rotate it clockwise to
horizontal position.
It is not possible to remove the key when
set to horizontal position.
Turn on the laser unit (see section 6.8
"Switching on the Unit", page 50). During
start-up, the Limax
a self-test for approx. 10 – 30 seconds.
The blue start-up screen shows the
KLS Martin logo, the Limax
the software version number 0.2.
Check the start-up screen for potential
problems or error messages.
The screen must comply with the screenshot
on the left.
Once the system test has been completed
successfully, the Standard program will be
displayed.
Operating Instructions
®
120 routinely performs
®
120 logo and
®
120