6.2

Connecting the Laser Fiber

NOTICE
The laser features a coded SMA (SMAplus) socket which ensures that laser activation is
possible only if a laser fiber with an SMAplus plug has been connected.
Remove the laser fibers and applicators permitted for use with the Limax
chapter 9.2 "Accessories", page 79) from their packaging and connect them to the
®
Limax
120. When using sterile disposable products, check the packaging for integrity
and observe the expiration date!
Before connecting the supply fiber of a laser focusing handpiece, the following visual
inspection steps must be carried out:
— Check supply fiber for potential damage over its entire length.
— Remove protective cap from the SMAplus plug of the supply fiber.
— Check coupling-in side of the SMAplus plug for potential damage and soiling.
— Inspect the optics (lens) of the laser focusing handpiece.
Observe the fiber-specific Instructions for Use and the instructions provided on the
applicator packaging.
Insert the SMAplus plug axially into the SMAplus connecting socket, then fully fingertighten
the nut (do not use tools!). Once the SMAplus plug has been connected correctly, the red
"Fiber missing" symbol disappears from the screen and "LASER STAND BY" is displayed in
green.
Check the fiber and the handpiece for proper working condition using the pilot laser light
(see section below).