Configuring the System
When PACE detection is set to Weak, the pacemaker signals are not easy to be detected.

9.2.5 Setting HR Calculation Lead

You can set the HR calculation to one of the 12 standard leads: І, П, Ш, aVR, aVL, aVF, V1, V2,
V3, V4, V5, or V6.

9.2.6 Selecting Auto Printing When Detecting Arrhythmia

When Auto printing when arrhythmia is selected, if Arrhythmia ECG data, including
Ventricular Tachycardia, 5>PVCS>=3, Paired PVCS, Bigeminy, Trigeminy, R ON T, single PVC
and Missed Beat, is detected during the sampling course, printing will be triggered automatically.

9.2.7 Setting Auto Upload When Stopping Sampling

When Auto upload when sampling stops! is selected, the date will be uploaded to ENS after
sampling is stopped.

9.2.8 Setting Background Grid

Select Background Grid, the grid on the background of the ECG sampling interface will be
displayed.
Deselect Background Grid, the grid on the background of the ECG sampling interface will not
be displayed.

9.2.9 Setting Anti-aliasing

Select Anti-aliasing, the system will automatically make the waveform smooth.
Deselect Anti-aliasing, the system will not make the waveform smooth.

9.2.10 Selecting QRS Voice

If you select QRS Voice, there will be a beep when an R wave is detected.
- 114 -