Scan - Thermo Scientific GENESYS 30 User Manual

Spectrophotometer
Table of Contents

Advertisement

Scan

Message bar - look here for
instructions and reports
Measurement mode
selection and indicator tool
Scan upper wavelength limit
field
Scan lower wavelength limit
field
Toggle cursor on/off tool
(only active when a scan is
displayed)
Thermo Scientific
Zoom tools
Save tool
Clear tool
To record a scan
1. Select the data mode, ABS or %T.
2. Enter the lower and upper scan limits (minimum scan range is 100 nm).
3. Place a cuvette with a blank in the measurement position, close the lid and press
4. Remove the blank cuvette, place the sample cuvette in the measurement position, close
the lid and press
The scan plot y-axis autoscales as the data is collected.
The current measurement wavelength and ABS/%T are displayed as the scan is recorded.
Wavelength and ABS/%T
values at cursor position
.
Baseline status
indicator
.
GENESYS 30 User Guide
11