Thermo Scientific GENESYS 30 User Manual page 16

Spectrophotometer
Table of Contents

Advertisement

14
GENESYS 30 User Guide
4. Enter the concentrations for all your standards; navigate to
Follow the directions displayed in the message bar.
5. Place a cuvette with a blank solution in the measurement position and press
6. Present and measure your standards in order as instructed in the message bar.
Tip If you wish to re-measure a standard, navigate back to the Conc field for that
standard and follow the directions in the message bar.
After you measure the last standard, the focus will move to
accept the calibration data.
Tip You can save your method at any time by navigating to and selecting
have measured all the standards listed in the table you may save the method as
calibrated or uncalibrated. If any of the standards has not yet been measured you can
only save the method as an uncalibrated method.
The screen moves to the Quant working screen.
Tip Select
to go back to the calibration screen and re-measure a standard if the
graph shows that your data does not represent a good enough fit.
Follow the instructions in the message bar to measure your samples.
The application supports a data table with entries for up to ten samples. Samples are labeled
S1 through S10. Once the data table is full you must clear it before you can measure
additional samples.
and press
.
.
. Press
to
. If you