Draining Vapor

– before it is moved around if there is a risk that it will
be tilted by more than 45°,
– when it is not going to be used for more than one
month.
Anaesthetic agent vapours must not be inhaled.
Danger to health.

Vapor with key-indexed filling system

1 Switch off Vapor = turn handwheel to »0« –
»0« button engaged.
Use correct bottle of specific anaesthetic agent.
Screw suitable adaptor firmly onto bottle.
1 Switch fill/drain valve to »0«.
2 Unscrew locking screw to the stop, as the seal may
be damaged otherwise when inserting the adaptor.
3 Pull out sealing block.