Technical Data
Filling volume for anaesthetic agent
Consumption of anaesthetic agent [mL/hour]
Rough formula for running time [hours] =
(for 150 mL anaesthetic agent)
Pressure range
Loss of anaesthetic agent when
switched off
Maximum angle of tilt, filled
Weight, filled
Dimensions Vapor 19.n
with key-indexed filling
system (measured in mm)
Measurements in brackets apply to
Vapor 19.n with filling spout
Vapor conforms to relevant Standards
Key-indexed filling system
________________
Standards marked demand a key-indexed filling system
*
28
‫.ﻣراﺟﻌﮫ ﻓرﻣﺎﯾﯾد‬WWW.ELM-FAN.IR‫ﺑرای داﻧﻠود ﻣوارد ﺑﯾﺷﺗر ﺑﮫ ﺳﺎﯾت‬
about 200 mL with dry wick
about 140 mL with moist wick
about 135 mL between minimum and maximum mark
~3 x fresh gas flow [L/min] x concentration [vol.%]
fresh gas flow [L/min] concentration [vol.%]
Example: fresh gas flow = 2 L/min, concentration = 1.5 vol.%
–100 mbar to 200 mbar
<0.25 mL liquid/24 hours
45° (with fixed connection on anaesthetic machine)
6.5 kg to 8.5 kg depending on design
Vol.
ø 106
116 (106)
Vapor 19.3: a = 12 mm
Vapor 19.2: a = 8 mm
Vapor 19.1: a = nicht vorhanden
DIN 13252*
ISO 5358
BS 4272
SN 057600
CSA Z168.3*
ASTM F1161
prEN 740 (draft)*
ISO 5360
prEN 1280
CSA Z168.4
To download more items, go to ELM-FAN.IR
50
Running time = 16.5 hours
0
%
not available
143 (135)