NS 5.14 Firmware
5. Refer to "Network Mode" on page 4-5 to enter values for the
Network menu.
6. The Channel 1 Connection menu displays the Ethernet ports TCP
connection settings, including the connection port. Refer to Figure
4–4 below.
Figure 4–4 Connection window (Web Manager)
7. Refer to "Channel Connection" on page 4-7 to enter values for the
Channel 1 Connection menu.
In the "Channel 1 Serial Settings" menu, do not change the
following parameters: Protocol (RS232), Flow Control (None),
Baud Rate (230400), Data Bits (8), Parity (None), and Stop Bits
(1).
8. When finished, save and apply the configuration changes by clicking
Apply Settings under the menu options. This will save the settings
and reboot the Ethernet port. This function will take approximately
60 seconds.
Clicking OK on each page does not change the configuration on
the Ethernet port, but simply confirms which changes to use.
Selecting "Apply Changes" will make the changes permanent and
initiate a reboot.
All remaining parameters should be left at default values unless
the user is knowledgeable in configuring the Ethernet.
4
4
4900002262 rev. B 3-23-21