4. Set the Start Validation parameter to 1 and then press the * key
followed by the # and the 1 key to allow the validation to start. Allow
the validation sequence to complete.
5. Press the # key followed by the 2 key (to enter Mode 2) and then
press the * key twice.
6. View the newly calculated RATA values, and if desired, set Update
RATA to 1 to accept the changes. Leave the setting at 0 to reject
the changes and start over.
7. Re-run steps 2 through 4.
8. Press the # key followed by the 9 key (to enter Mode 9) to verify
the validation results and confirm that new RATA values are
operating correctly.
9. Press the # key followed by the 1 key (to enter Mode 1) to return
to Normal Mode.
Automatic single or dual validation
Either one or two standards can be used for automatic single or dual validation;
Validation 1 and/or Validation 2, which are automatically introduced and
controlled by the analyzer. The validation is automatically initiated based on the
hour of the day and is controlled by the Val Start Time, Daily Validation and
Val Interval parameters. Validation can also be initiated remotely using the
Start Validation parameter or the Start Validation digital input.
1. Configure the Operator Parameter section with the Operator
Password set to 0 and one parameter.
a. Configure Update RATA by setting Operator Parameter01 to
225. Other parameters can be added, if desired.
2. Press the # key followed by the 9 key (to enter Mode 9) to verify
the validation results.
3. Press the # key followed by the 2 key (to enter Mode 2) and then
press the * key.
4. View the newly calculated RATA values, and if desired, set Update
RATA to 1 to accept the changes. Leave the setting at 0 to reject
the changes and start over.
5. Press the # key followed by the 1 key (to enter Mode 1) to return
to Normal Mode.