Custom Precision

The Custom Precision parameter sets the number of viewable digits to the
right of the decimal point. The total number of digits the analyzer can display
at any one time is six. Therefore, when the size of the value plus the Custom
Precision exceeds six, the number of digits to the right of the decimal point will
be reduced accordingly.

Daily Validation

The Daily Validation parameter enables or disables the time of day
autovalidation feature. When enabled, an autovalidation cycle is initiated every
'X' day (where 'X' is defined by Val Interval) at the time of day established by
Val Start Time. Enter 0 to turn the feature off or 1 to turn the feature on.

Dew Point Method

The Dew Point Method parameter sets the type of industry standard dew
point calculation to be performed when Calculate Dew Point is enabled. Enter
0 for ISO 18453:2006, 1 for ASTM 1142-95 Eq. (1), 2 for ASTM 1142-95 Eq.
(2), and 3 for the Arden Buck Method.
Operator's Manual
<SET PARAMETER MODE>
Concentration Unit
0
0:ppm 1:lbs 2:% 3:mg
<SET PARAMETER MODE>
Custom Precision
2
Enter a value
<SET PARAMETER MODE>
Daily Validation
0
0:Disable 1:Enable
Operating the Analyzer