Start Validation

The Start Validation parameter initiates the validation cycle. After the cycle
begins, this parameter automatically reverts to 0.

Temperature Unit

The Temperature Unit parameter designates the display units for the
measured cell temperature. There are two choices: 0 for degrees Celsius and
1 for Fahrenheit. The default value is the standard unit of measurement in the
region the analyzer is being used.

Update RATA

The Update RATA parameter is used to update the RATA Multiplier and
RATA Offset parameters with the latest automatically calculated values. Each
time an automatic, semi-automatic, manual Mode 7 or manual Mode 8
validation is completed, a new RATA Multiplier and RATA Offset will be
calculated. This parameter displays the current RATA Multiplier and RATA
Offset in the left column of the display as defined in their respective parameter
descriptions. In the right column, the newly calculated values are displayed. To
accept and use the newly calculated values, enter 1. These values will be
invoked when pressing the MODE button while exiting Mode 2.
Refer to the section called "Adjusting Analyzer Reading to
Match Specific Standard(s)" on page 2-35 section for more
information.
Operator's Manual
<SET PARAMETER MODE>
Start Validation
1
1:Start
<SET PARAMETER MODE>
Temperature Unit
0
0:C 1:F
<SET PARAMETER MODE>
Mult: 1.00 New: 1.00
Ofst: 0.00 New: 0.00
0
1: Update RATA
Operating the Analyzer
2
31