NS 5.14 Firmware
Table 2-1 Typical values for parameter setpoints (Continued)
Parameter
RATA
(Relative Accuracy Test
Audit)
RATA Multiplier
RATA Offset
Update RATA
Peak Tracking
Keypad Watchdog
Set Time - Hour
Set Time - Minute
Set Time - Day
Set Time - Month
Set Time - Year
General Alarm DO
DO Alarm Setup
Low Alarm Setpoint
High Alarm Setpoint
AO 4-20 mA Test
4-20 mA Alarm Action
–
2
12
Setting
0 or 1
Default: 0
-1.E+06 to 1.E+06
Default 1.0
-1.E+06 to 1.E+06
Default 0.0
0 or 1
Default: 0
0, 1 or 2
Default: 0
< 5:Off, >=5:Secs
Default = 10
0 to 23
Default: 0
0 to 59
Default: 0
1 to 31
Default: 1
1 to 12
Default: 1
2006 to 2144
Default: 2017
0, 1, or 2
Default = 0
4.29E+09
Default = 8192
-1.E+06 to 1.E+06
Default = -10000
-1.E+06 to 1.E+06
Default = 10000
0.0 to 100.0
Default: 0.0
0, 1, 2 or 3
Default: 0
Function
Enables or disables adjustment
factors.
Slope adjustment factor.
Offset adjustment factor.
Updates RATA Multiplier and
Offset to automatically
calculated values.
Sets peak tracking capability to
off, on or reset.
The time-out on the Mode
screen before the analyzer
automatically reverts to Normal
Mode.
Sets the current hour.
Sets the current minute.
Sets the present day.
Sets the present month.
Sets the present year.
Sets the alarm mode (latching
or non-latching) and resets the
general alarm once activated.
Sets the functionality of the
secondary digital output.
Sets the low alarm threshold
for moving average.
Sets the high alarm threshold
for moving average.
Sets the 4-20 mA output to a
percentage of full scale.
Sets the current loop state
upon alarm condition.
4900002262 rev. B 3-23-21