Dry Temp (C): Measured temperature (0.0 for non-differential
units) of the gas sample from the last scrubber cycle in degrees
Celsius.
•
Ethane: Sets the mole fraction of ethane in the dry gas mixture used
when calculating the dew point temperature per ISO 18453:2006.
The default value (0.08360) corresponds to the natural gas mixture
NG3.
•
FitResidue: Value ranging from 0 to 1 indicating how well the
measured spectrum fits the referenced spectrum, where 1 represents
a perfect match.
•
Fitting Ratio: Ratio of the measured spectrum versus the Reference
1 or Validation spectrum.
•
Fitting Ratio 2: Ratio of the measured spectrum versus the
Reference 2 spectrum, where a value of 0 indicates that the
Reference 2 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio 3: Ratio of the measured spectrum versus the
Reference 3 spectrum, where a value of 0 indicates that the
Reference 3 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio 4: Ratio of the measured spectrum versus the
Reference 4 spectrum, where a value of 0 indicates that the
Reference 4 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio 5: Ratio of the measured spectrum versus the
Reference 5 spectrum, where a value of 0 indicates that the
Reference 5 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio Dry: Ratio of the measured spectrum versus the
Reference Dry spectrum, where a value of 0 indicates that the
Reference Dry spectrum was not used in the calculation of
concentration.
•
Fitting Ratio Dry-1: Ratio of the measured spectrum versus the
Reference Dry spectrum shifted by 1 index value, where a value of 0
indicates that the Reference Dry spectrum shifted by 1 index value
was not used in the calculation of concentration.
•
General Alarm DO: Sets the operation of the general alarm relay
digital output when a general fault alarm occurs and allows it to be
reset if needed.
•
grn ConvFactor 07: Sets a custom conversion factor when the
parameter Concentration Unit = 7 (grains/100 scf) and this value is
greater than 0.0. The default value conversion factor is used when it
is equal to 0.0.