Thermo Scientific Partisol 2025i Instruction Manual page 107

Sequential air sampler
Table of Contents

Advertisement

Thermo Fisher Scientific
If the Sampler is in the Audit mode, press
the Main Menu, scroll to Audit and Calibration and press
display the Audit and Calibration menu, with the cursor at Audit
mode, press
Mode screen, press
confirm the change.
2. On the Main screen, confirm that the date and time are correct.
If the date or time is not correct, in the Main Menu, scroll to
Instrument Setup and press
scroll to Date/Time and press
screen. In the Date and Time screen, press
mode and follow the screen directions to edit the date and/or time.
When the changes are complete, press
to return to the Instrument Setup menu.
3. Confirm that the Site ID is correct. In the Main Menu, scroll to
Instrument Setup and press
menu. In the Instrument Setup menu, scroll to Site ID, press
and confirm that the Site ID is correct.
If the Site ID is not correct, scroll to the incorrect Site ID, press
to display the alphanumeric entry screen, enter the correct Site ID
description, and save it.
For information on the Site ID, refer to
For information on how to use the alphanumeric entry screen, refer to
"Alphanumeric Entry
4. Press
repeatedly to return to the Main Menu. In the Main Menu
with the cursor at Sample Setup, press
Setup menu. The Sample Setup menu allows you to choose modes such
as:
Auto Populate which calculates the default start time and date once
sample 01 is programmed
Apply EPA Times which is a shortcut key for midnight-to-
midnight, 24-hour continuous sampling
Apply Default Times which automatically applies the default time
settings.
To apply EPA default conditions (midnight-to-midnight, continuous, 24-
hour sampling) scroll to Apply EPA Times in the Sample Setup menu,
to display the Audit Mode screen. In the Audit
to exit Audit mode and press
to display Instrument Setup menu,
to display the Date and Time
to display the Instrument Setup
"Site
Screen" on page 9-4.
Partisol 2025i and Partisol 2025i-D Instruction Manual
Sampler Operation
Programming a Sample Run
repeatedly to display
to
to
to access the Edit
to save the changes. Press
ID" on page 9-37.
to display the Sample
,
7-13