Inspection Of The I.i. Support; Line Voltage And Internal Line Impedance - Siemens LITHOSTAR Multiline SP Start-Up Instructions

Table of Contents

Advertisement

General Start-Up

Inspection of the I.I. support

Remove both covers from the image intensifier (
Check the image intensifier support for damage.
Replace the image intensifier if defects are found.
Reattach the covers if no defects are found.

Line Voltage and Internal Line Impedance

CAUTION
Remove the cover from the knife fuses F1, F2 and F3 in M16 of the generator and
remove the fuses using the special gripper.
Connect a multimeter between two of the phases at the lower connections of the fuse
holders for F1, F2 and F3 (X2169-11, POLYDOROS LX / X2206-10, POLYDOROS SX).
Switch the system breaker ON.
Use the multimeter to measure the voltage, phase to phase, consecutively.
p
Record the voltage measured in the Multiline test report.
Switch the system breaker OFF.
Measure the internal line impedance at M16.L1, L2, L3 using the internal line impedance
meter (phase-to-phase) as follows:
Connect the internal line impedance meter between 2 phases respectively of L1, L2 and
L3.
Switch the system breaker ON.
Perform the measurement.
Siemens AG
Medical Solutions
The LITHOSTAR Multiline system can be operated only at a line
voltage of 400 V ± 10% and 50/60 Hz. A system pretransformer
must be used for all other line voltages.
RXL2-120.815.01
Rev. 02
Fig. 1
/Fig.1).
Page 1 of 4
02.05
CS PS 24
2 - 1
2
2