6 - 10
If the dose rate is acceptable:
- Connect the service PC to the Multiline.
- Call up "Parameter" in the Adjustment menu.
- Click on "Get from Unit."
- Select the parameter :" am 7 TV-Iris" in the Parameter Groups menu.
- Transfer the value appearing at "FLUORO Zoom 0" to "FLUORO Zoom 1" and
"FLUORO Zoom 2".
Dose rate setting for ZOOM 3:
NOTICE
- The dose rate for FLUORO Zoom 3 must be set separately.
Multiline without grid:
Multiline with grid:
Note the factor for the chamber.
If you are using the Dali -/Nomex-chamber for the measurement, the dose rate value
must be increased by 6%.
Leave the dose measurement chamber attached to the I.I.
Connect the service PC to the Multiline and login.
Release fluoro.
Call up "Parameter" in the Adjustment menu.
Click on "Get from Unit".
Select the parameter : "am 7 TV-Iris" in the Parameter Groups menu.
Select the "FLUORO" " Zoom 3" parameter.
In the "Value Actual" field, reduce or increase the zoom value.
- If the dose is too high, increase the value
After completing the change, click "Set Value".
The changed value is marked in the parameters window.
(It is not yet transferred to the unit.)
Click on "Put to Unit". The message "Component ... to continue" appears.
Click "OK" after the automatic reset.
Switch fluoro on if the dose value is acceptable.
End the adjustment, otherwise repeat it.
NOTE
LITHOSTAR Multiline
System manual
Zoom3 is not set like Zoom 0. The videomed DI service software
does not perform a nominal-actual comparison.
430 nGy/s
670 nGy/s
After each Reset of the Multiline, the antiisowatt curve must be
selected on the operating PC.
RXL2-120.815.01
Rev. 02
Checking the Dose Rate
33 cm I.I. Zoom 3
49 µR/s
77 µR/s
Page 10 of 18
02.05
CS PS 24
Siemens AG
Medical Solutions