Checking the Dose Rate
Message line: Start fluoroscopy with ON
Message line: Turn D100 R1 in the camera adapter to the right.
Set the adjustment in the camera adapter to +1 (+1/4 EP) with R1 on board D100.
If this range is not adequate, change over to switch S2 on board D100 in the camera
adapter and repeat the adjustment.
Remove the measurement chamber.
Leave 1.2 mm Cu filtration and exit the window with OK; the window PDA Preadjust
appears.
With fluoro ON, check whether the value 0 +1 is attained in the field "Difference of
expected to actual value".
Otherwise, adjust potentiometer R1 on D100 until the adjustment is successful.
Siemens AG
Medical Solutions
7
RXL2-120.815.01
Rev. 02
S2
R1
negative value −−> CCW
Page 15 of 18
02.05
CS PS 24
6 - 15