8 - 2
Attach the generator front panel, connect the protective conductor and secure it with
screws (POLYDOROS LX only).
Install the uro insert and check all unit settings.
Remove the uro insert, move the shock head into the park position and check all unit
settings.
Move the shock head into working position with the uro insert removed and check all unit
end settings.
Activate the anti-collision switch by pressing on the tube unit housing. Unit movements
are blocked.
Start unit movement in the patient direction. Activate the anti-collision switch of the tube
unit housing at the same time and hold it in this position. The unit movement must be
stopped.
Activate one of the two hazard switches.
Move the unit out of the collision zone.
Move the unit and at the same time activate the lateral switching strips on the shock
wave head. The unit movements must be blocked.
Attach the cassette holder.
Raise the table.
Push the cassette holder upward during the movement. This must block the drive and
the LED's in the danger buttons must light up.
Push the EMERGENCY STOP button during the unit movement. All movements must be
switched off.
Clean the system.
For further information (cleaning/disinfecting), refer to the Operating Instructions.
The manufacturer of this product requires information and the legislator demands proofs
that a product delivered free of defects from the factory continues to possess the
required and certified quality properties on installation and start-up.
It is therefore absolutely necessary that the installation report with the installation and
start-up data is sent without delay after completion of the work to the address stated on
the installation report.
You will find the report in the system folder.
NOTICE
NOTE
LITHOSTAR Multiline
System manual
After completing all adjustment steps and test measurements, the
service engineer must confirm with the date and signature that all
values have been set correctly and have been recorded in the
Test Certificates (unit, generator and IQ Test Certificates).
From serial number 7000 the laser light localizer can be converted
to battery operation (see instructions RXL2-120.033.03... ).
RXL2-120.815.01
Rev. 02
Page 2 of 2
02.05
CS PS 24
Final Steps