4 - 4
Confirm the message "Please close the collimator" with OK.
Select "Fluoroscopy" in the Adjustment menu.
Switch on fluoroscopy (i.e. click "on" using the mouse). After approximately 10 minutes,
fluoroscopy switches off automatically.
Click OK.
Switch on "Fluoroscopy" in the Pre-Heating field (click ON with the mouse).
Fluoroscopy switches off automatically.
Click OK after the fluoro has switched off.
Switch on "Correction Exposure" in the Learn Filament field (i.e. click ON using the
mouse). The correction is determined for Load 10 %.
Release another exposure after the correction. The correction for Load 60 %
(large focus) and 80 % (small focus) is performed.
Click OK.
Release an exposure in the Learn Filament Pushcurrent field (i.e. click ON with the
mouse).
Continue to release exposures until the message: "Release next exposure" disappears.
Click OK.
The query appears: Next focus (click cancel if second focus is set already). If you need to
perform the adjustment for the second focus, click OK in the ADJ menu.
Select warm-up and repeat the procedure.
Click OK after fluoro switches off.
Switch on an exposure in the Learn Filament Correction field (click ON using the mouse).
The correction is determined for Load 10%.
Release another exposure after the correction. The correction is determined for Load
60 % (large focus).
Click OK.
Release an exposure in the Learn Filament Pushcurrent field (i.e. click ON using the
mouse).
Release additional exposures until the message: "Please wait...exposure" no longer
appears.
Click OK.
Click CTRL + X in the System Quit menu.
Reset switch S3/X44.
Confirm the subsequent message "Set the S3/X44 switch ..." with OK. The green LED
V90 on board D100 goes out (refer to Fig. 2).
Click OK.
Click "Backup to Disk" in the Data menu.
Insert the Site data floppy.
Click OK.
LITHOSTAR Multiline
System manual
Start-Up of Multiline System
RXL2-120.815.01
Rev. 02
02.05
Page 4 of 8
CS PS 24
Siemens AG
Medical Solutions