General Start-Up
Fig. 2

Generator Start-Up

Switch off circuit breakers F7, F8 and F9 in M16 of the Multiline control cabinet.
Switch the system breaker
ON:
NOTE
Switch the system ON; The generator initializes; the 7-segment display on D100
Use S2 on D160 to switch the generator OFF (test of blocking). It may not be possible to
switch on the generator from the control console.
Set sliding switch S2 on D160 position 1.
System contactor OFF.
Switch on circuit breakers F7, F8, F9 in the M16 control cabinet.
Switch S1 on shock wave cabinet N12 ON.
Withdraw plug from pump M15 in N12 (pump must not run dry).
Switch the system ON.
The message "No contact to shock wave system" appears on the PC monitor.
NOTE
Siemens AG
Medical Solutions
The red LED V23 and the lamps H1 and H2 must light up
on power-on circuit board D160.
In the event of a fault check the fuses for transformer T1 (below
D160) (X2169-11, 12 and 19 and X2206-11, 12 and 19).
Due to changes in the monitoring of the shock wave system, the
shock wave cabinet must be switched on when switching on the
system. The shock wave cabinet can be switched off, when the
system is ready for use.
RXL2-120.815.01
Rev. 02
displays "b" (meaning "ready").
Page 3 of 4
02.05
CS PS 24
2 - 3
2
LITHOSTAR Multiline
System manual