Filling The Cooling System - Siemens LITHOSTAR Multiline SP Start-Up Instructions

Table of Contents

Advertisement

Shock Wave System

Filling the Cooling System

NOTE
Move the shock wave head into the therapy position.
Disconnect the plug to the M15 in the N12 pump (the pump must not run without water).
Measure the room temperature.
Connect the multimeter at test point X2.2 of board W11 in slot W11A and connect the 0 V
lead to the D1 board at test point X3.
Check whether the voltage measured matches the room temperature value from the
following table.
Voltage tolerance: ± 30 mV; adjustable using R13 on the W11 board.
°C
Volt
° F
°C
Volt
° F
Turn N12 off with S1.
Reconnect the connector to the M15 pump.
Fill the water container with 4 liters of sterile water (refer to Speed Info RX 17/96).
Install the water container in the cabinet on the transport bracket.
NOTE
Ensure that the vent valve on the filter is set to open. If not, set it to open by holding the
lower brass screw in place with a crescent wrench.
Open the red cap on the vent valve.
Open the quick couplings and pump outlet (located on the right hand-side behind the
pump).
Turn on N12 using switch S1 until water runs out of the fast couplings.
Turn off N12 using switch S1.
Connect the quick couplings and the pump.
Turn on N12 using switch S1; the cooling circuit is filled and ventilated.
NOTE
Siemens AG
Medical Solutions
Due to changes in the monitoring of the shock wave system, the
shock wave cabinet must be switched on when switching on the
system. The shock wave cabinet can be switched off, when the
system is ready for use.
18° C
19° C
20° C
4.36
4.53
4.69
64° F
66° F
68° F
27° C
28° C
29° C
5.82
5.98
6.14
81° F
82° F
84° F
The container should not be closed to allow adequate ventilation.
Ensure that the cooling circuit does not contain air bubbles.
21° C
22° C
4.85
5.01
70° F
72° F
30° C
31° C
6.31
6.47
86° F
88° F
RXL2-120.815.01
Rev. 02
02.05
3
23° C
24° C
25° C
5.17
5.33
5.50
73° F
75° F
77° F
32° C
33° C
34° C
6.63
6.79
6.95
90° F
91° F
93° F
Page 1 of 6
LITHOSTAR Multiline
CS PS 24
3 - 1
3
26° C
5.66
79° F
35° C
7.11
95° F
System manual

Advertisement

Table of Contents
loading

Table of Contents