Protective Measures; Preliminary Steps - Siemens LITHOSTAR Multiline SP Start-Up Instructions

Table of Contents

Advertisement

Prerequisites / General Information

Protective Measures

Adhere to the protective measures contained of the LITHOSTAR Multiline System
binder.
NOTICE
WARNING
Set the main system switch to OFF to switch off power to all components (generator and
connected equipment).
Switch OFF (S1) SS on the D 100 board (Fig. 1, no inverter power) to avoid inadvertent
release of high voltage or radiation.
Install or remove assemblies only with the generator switched off and according to ESD
guidelines.

Preliminary Steps

Set switch (S1) on board D 100 to SS OFF (Fig. 1).
Siemens AG
Medical Solutions
Tests and adjustments performed with radiation ON are identified
by the radiation warning symbol
Radiation protection must be worn during these types of adjust-
ments (refer to ARTD part 2).
Voltage is still present in transformer T1 and power-on circuit
D160 even after the generator is switched off (see wiring dia-
grams X2206-10, X2169-10).
Following generator switch-off a voltage of approximately 600 V
DC is still present for the converter! LEDs V35 and V36 on D110
(Fig. 1) indicate this (X2206-16, X2169-10) and LED V89 (X2206-19,
X2168-019) lights up.
The voltage drops to 0 V in approx. 1
when the voltage reaches approximately 30 V.
RXL2-120.815.01
Rev. 02
.
1
/
minutes; the LED's go out
2
Page 3 of 6
02.05
CS PS 24
1 - 3
1
1
LITHOSTAR Multiline
System manual