Siemens LITHOSTAR Multiline SP Start-Up Instructions page 37

Table of Contents

Advertisement

Checking the Dose Rate
Record the measurement value (approx. 3.1 mA).
Multiline without grid:
Multiline with grid:
Exit the dose rate adjustment field with "OK" after the dose rate has been reached.
Enter "150 mV" in the B-Signal value field.
Confirm the B-signal value with OK.
Click "Fluoroscopy on" in the CCD Brightness Adjust window.
Ignore the numerical value and click "Fluoroscopy off".
Click Cancel.
Confirm the message "Iris setting ..." with OK.
Select Dose rate in the Adjustment menu.
The message "IImaximum ... " appears.
Select the I.I. full field of view and open the collimator until it matches the size of the
measurement chamber.
Confirm the message "IImaximum ... " with OK.
Click Fluoroscopy on.
Change the mA value so that the dose rate desired appears on the dosimeter.
Record the measurement value (approx. 3.1 mA).
Multiline without grid:
Multiline with grid:
Enter "300 mV" in the B-Signal value field.
Click OK.
Wait for the data to be transferred to the generator (this may take some time).
Click "Fluoroscopy on" in the CCD Brightness Adjust field if the value 0 + 1 is not
reached.
Exit the XCS software.
- Switch fluoroscopy ON.
- Read out the nominal/actual value. Record values and mathematical signs if there are
deviations (refer to RXL2-120.034.02..).
- Switch fluoroscopy OFF.
- Connect the service PC to the Multiline if there are deviations.
- Call up "Parameter" in the Adjustment menu.
- Click Get from Unit.
- Select the parameter: am 7 TV-Iris in "Parameter Groups".
- Select "TV-Iris format" under Parameter.
Siemens AG
Medical Solutions
Dose rate
261 nGy/s
30
433 nGy/s
49
Dose rate
261 nGy/s
30
433 nGy/s
49
RXL2-120.815.01
Rev. 02
02.05
B-Signal
µ
R/s
300 mV
µ
R/s
300 mV
B-Signal
µ
R/s
300 mV
µ
R/s
300 mV
Page 7 of 18
LITHOSTAR Multiline
CS PS 24
6 - 7
System manual