Siemens LITHOSTAR Multiline SP Start-Up Instructions page 12

Table of Contents

Advertisement

2 - 2
Switch the system breaker OFF.
Remove the internal line impedance meter and test the other phases.
p
Record the measured value in the table below or in the Multiline test report.
To reach full generator power, the measured internal power line impedance may not
exceed the following value.
U
400 V ± 10%
400 V ± 10%
400 V ± 10%
Switch the system breaker OFF.
Reinstall the knife fuses.
Reinstall the cover over the knife fuses.
LITHOSTAR Multiline
System manual
max. R
Line
Line
max. R
Line
max. R
for POLYDOROS LX 50
0.15 Ohm
for POLYDOROS SX 65
0.14 Ohm
POLYDOROS SX 80
Line
0.11 Ohm
RXL2-120.815.01
Rev. 02
02.05
General Start-Up
Measured
Page 2 of 4
CS PS 24
Siemens AG
Medical Solutions