Siemens LITHOSTAR Multiline SP Start-Up Instructions page 18

Table of Contents

Advertisement

3 - 4
Fig. 3
Fill the coupling bellows with water by activating the S1 key on board W12 (Fig. 1)
several times.
NOTICE
Remove any air that has collected in the coupling bellows using a syringe without a
needle.
Set switch S2 to the EMPTY COUPL. BELLOWS position (Fig. 1) (valve Y3 is open).
Connect the funnel to fill the water system (2/Fig. 2).
Fill the water tank so that the water level is between the marks on the upper sight glass
(1/Fig. 2).
NOTICE
Remove the funnel (release the quick-connector on the water tank) (2/Fig. 2).
Set the S2 switch on board W12 to the STANDARD MODE position (Fig. 1) (Y3 valve
open).
Fill the coupling bellows with water using the S1 button on W12 board (Fig. 1).
NOTICE
Remove air bubbles in the coupling bellows using a syringe (without a needle) (Fig. 3).
Set the S2 switch on board W12 to the EMPTY COUPLING BELLOWS position (Fig. 1)
(Y3 valve is open).
LITHOSTAR Multiline
System manual
The level in the water tank must not drop below the mid range of
the lower sight glass (3/Fig.2).
Fill the water tank only when the coupling bellows has completely
collapsed (a minimal amount of water remains in the coupling bel-
lows).
The level in the water tank must not drop below the mid range of
the lower sight glass (3/Fig. 2).
RXL2-120.815.01
Rev. 02
Shock Wave System
Page 4 of 6
02.05
CS PS 24
Siemens AG
Medical Solutions