EDAN SE-12 Series User Manual page 82

Electrocardiograph
Hide thumbs Also See for SE-12 Series:
Table of Contents

Advertisement

SE-12 Series Electrocardiograph User Manual
Message
U Disk / USB Printer / USB
Scanner / Reader
Message
Battery Weak
No Paper
Paper Error
Module Error
DEMO
Lead X off
Overload
HR Overrange!
Sys. Overrange!
Dia. Overrange!
Lanyard Off
Trans.Comm Err
Transducer Err
Incline Err
Great Interference
No Speed Signal
Cause
A U disk, a USB printer or a bar code reader / security reader / ID
card reader is connected to the USB interface.
Table 12-2 Error Messages and Causes for Exercise ECG
Cause
The battery is weak.
Recorder paper runs out or is not loaded.
When Paper Marker is set to Yes, the electrocardiograph
advances the recorder paper to the next black marker. If it
advances the paper for 300mm and cannot find the next black
marker, the hint Paper Error is displayed.
There is something wrong with the signal sample module.
The system is in the demonstration mode.
Electrodes fall off the patient or the patient cable falls off the unit.
The direct current offset voltage on an electrode is too high.
Heart rate exceeds the normal range.
Systolic blood pressure exceeds the normal range.
Diastolic blood pressure exceeds the normal range.
The safety stop (mushroom type) is revolved and the safety stop
(cord type) is pulled out.
The transducer and the main control board lines are not connected
well.
The transducer doesn't work well or is not connected well.
The incline motor doesn't work well or is not connected well.
There is great magnetic interference around the transducer.
The speed sensor doesn't work well or is not connected well.
- 76 -