EDAN SE-12 Series User Manual page 76

Electrocardiograph
Hide thumbs Also See for SE-12 Series:
Table of Contents

Advertisement

SE-12 Series Electrocardiograph User Manual
Operation Method:
1. Instruct the patient, attach the disposable electrodes to the patient, and then apply the stress
BP monitor to the patient. For details, please refer to Chapter 3 Operation Preparations.
2. Instruct the patient to lay on the bed, oberserve the ECG waveforms and the supine blood
pressure.
3. Select a protocol for the patient and configure the System Setup screen.
NOTE: You cannot change the settings of the System Setup screen after staring an
exercise test.
4. Press the Pretest key to begin the pretest phase.
5. Press the Exercise key to enter the exercise phase, and then observe the ECG waveforms,
heart rate, blood pressure, the patient's state and ST trend during the exercise test. If the stress
BP monitor is connected, the patient's blood pressure is measured once at a phase, and you
should print or save an ECG report after the main screen displays the BP value.
6. When the target value is reached, such as the target heart rate, press the Recovery key to
enter the recovery phase, and then instruct the patient to walk on the treadmill for 1 minute.
Observe the ECG waveforms, heart rate, blood pressure and the patient's state during the
recovery stage.
7. Instruct the patient to sit on the bed, and then observe the ECG waveforms and the blood
pressure for 6~8 minutes.
8. When the patient's heart rate resumes the normal value, press the Test end key to terminate
the exercise test. Press the PRINT/STOP key to print the final report. Select Trans on the
main screen of excersice screen to tranmist the report to the PC.
9. Take off the patient cable and electrodes, press the Pretest key to exit the test, and then
prepare for the next patient.
1. During the exercise test, ensure that such tests are supervised by properly trained
technician who meets competency requirements for exercise test supervision, fully
trained in cardiopulmonary resuscitation, and is supported by a physician skilled in
exercise testing or emergency medicine who is in close proximity for pretest
assessment or compliance that may raise.
2. Remind patients to take care not to fall down from the treadmill.
3. Press down the emergency stop switch of the treadmill before defibrillating to avoid
the hazard to the patient and the operator.
Operation Instructions for Exercise ECG
WARNING
- 70 -